Spider-80SG
General Data Acquisition Device with Strain Gage Measurement
The Spider-80SG is a high precision, general purpose data acquisition device featuring strain gage measurement. This device can be used in a variety of physical and measurement tests. The Spider-80SG can acquire data from a strain gage or a wide range of measurement quantities. A variety of general purpose and strain gage based sensors are supported. The Spider-80SG is built on the proven outstanding performance and reliability of Crystal Instruments’ DSP-based hardware platform. It features the same form factor as other Spider-80X front-ends from Crystal Instruments and can be configured into one measurement system with excellent compatibility and scalability. This gives the Spider-80SG capabilities to reliably acquire data from multiple sensors and for multiple measurement quantities simultaneously.
The Spider-80SG front-end connects to the network switch of a PC using an Ethernet connection. When used with a strain gage, it measures the strain based on arriving signals, measuring the changes in resistance when the strain gauge is stretched or compressed. It can also be used to measure Force, Torque, Pressure, Acceleration, Velocity and Displacement. The Spider-80SG can be configured to output an excitation voltage to power up the sensors and measure a range of measurement quantities. Pluggable front-ends provide the system with maximum flexibility of analog channel configuration, making it ideal for a range of measurement tests under various circumstances. The modular design allows configurations with 8 input channels of isolated analog inputs and one analog output.
With Crystal Instruments’ unique Ethernet based time synchronization technology, multiple Spider-80SG front-ends can be chained together to construct a system with higher number of input channels. High channel systems scale up to 512 channels.
Each Spider-80SG front-end has its own mass storage media that houses the operating software and stores measurement data. This truly distributed system guarantees data recording at full speed without being subject to network speed limitations.
Compatible Measurement Quantities & Suggested Sensors
Multiple sensors of various measurement quantities are supported with the Spider-80SG. The following list gives an overview of measurement quantities with sample sensors that are supported. The actual list of compatible sensors contains much more than specified here.
Acceleration – Dytran 7603B, 7503, 7523A2, Endevco 7264C, Kistler Type 8395A, DTS 6DX PRO series
Force – Omega LCM 901, Futek FFP350
Torque – Omega TQ-130 Series, Futek TDD400, Futek TRS300, Futek TAT200, TAT420
Pressure – Omega PX309 series, Measurement Spec EB100, Futek PMP927
Angular Velocity – DTS ARS Pro-300, ARS Pro-1500, ARS Pro-8K, ARS Pro-18k
Displacement – Omega E2E-3DC Series,
Magnetic Field – Analog Devices AD22151
Sound Pressure – InvenSense ICS40618
Strain – Configurable as Quarter, Half or Full Bridge
Hardware & Software Compatibility Specifications
The Spider-80SG can be used with the following devices:
Modes of Operation:
Standalone mode as a General Purpose Data acquisition device
In conjunction with Spider-80X for General purpose Data Acquisition
In conjunction with Spider-80X for general data acquisition during vibration control.
Product Specifications:
Input Channels
8 channels per front-end, expandable to 512 channels in a system
Connector Type
7-pin LEMO
Input Range
±5 mV, ±10 mV, ±50 mV, ±100 mV, ±10 V
Sampling Rate Per Channel
0.48 Hz to 256 kHz, with 58 stages
Shunt Calibration
Internal 100 K Ω (0.1%, 25ppm/c)
Excitation Sense
Local and remote sensing
Zero Suppression/Auto Balancing/Offset Nulling
Strain Functions
Quarter-120, Quarter-350, Half bridge, Full bridge
Bridge Completion
120 Ω: 0.05%, 5ppm/c
350 Ω: 0.05%, 5ppm/c
Back Half resistor: 10 K/10 K, 0.1% (or 0.02%), 2 ppm/c
Excitation Voltage
±2.5 V, ±5 V, ±10 V
Current: 30 mA max/channel